12 Flow Cytometry Terms And Definitions Most Scientists Get Wrong

The most important part of executing a flow cytometry experiment correctly is actually understanding what you are doing. This means you must understand the terms and definitions that are critical to the field of flow cytometry.

As a scientist, you should not just place your faith in a specialized technician. You should not blindly agree with the data you see in front of you without you yourself knowing what ‘logicle scaling’ means, what ‘differential pressure’ means, what happens when you change the differential pressure on an instrument, and so on.

You must also be able to communicate your methodologies and results intelligently. This means, for example, knowing the difference between flow cytometry, flow cytometry cell sorting, and FACS analysis: the latter merely being a trademarked term “owned” by a flow cytometry company.

Top 12 Most Commonly Misunderstood Flow Cytometry Terms

To help resolve this confusion, we have worked with our 700+ Mastery Class members to compile a list of the top 12 most commonly unknown or commonly misunderstood flow cytometry terms here…

1. Autofluorescence

Autofluorescence is the term given to describe the natural fluorescence that occurs in cells. The common compounds that give rise to this fluorescence signal include cyclic ring compounds like NAD(P)H, Collagen, and Riboflavin, as well as aromatic amino acids including tyrosine, tryptophan, phenylalanine. These compounds absorb in UV to Blue range (355-488 nm), and emit in the Blue to Green range (350-550 nm).

The consequence of this autofluorescence is the loss of signal resolution in these light ranges and a decrease in signal sensitivity. Autofluorescence typically increases with cell size. Larger cells have more autofluorescence than small cells due to the simple fact that the larger cells often contain more autofluorescent compounds.

2. Logicle Scaling

Logicle scaling is an implementation of biexponential scaling published by the Herzenberg lab at Stanford. The biexponential scale is a combination of linear and log scaling on a single axis using an arcsine function as its backbone.

The “logicle” implementation of biexponential was implemented in many popular software packages like FACSDiva and FlowJo. Other types of biexponential scaling exist, including Hyperlog.

Biexponential scales are more generally referred to as hybrid scales and include other variations like lin/log or log with negative. More information on logicle scaling can be found here: Parks DR et al. A new “Logicle” display method.

3. Fluorescence Minus One, or FMO Control

The Fluorescence Minus One control, or FMO control is a type of control used to properly interpret flow cytometry data.  It is used to identify and gate cells in the context of data spread due to the multiple fluorochromes in a given panel.

An FMO control contains all the fluorochromes in a panel, except for the one that is being measured. For example, in the 4-color antibody panel, there would be four separate FMO controls, as shown in the table below. The FMO control ensures that any spread of the fluorochromes into the channel of interest is properly identified.

Figure_1

4. Sheath Fluid

Sheath fluid is the solution that runs in a flow cytometer.  Once the sheath fluid is running at laminar flow, the cells are injected into the center of the stream, at a slightly higher pressure. The principles of hydrodynamic focusing cause the cells to align, single file in the direction of flow.

Depending on experimental needs, different formulations of sheath fluid can be used. Many labs purchase pre-mixed phosphate-buffered saline, while other labs use their own Hepes-buffered saline. The latter is particularly useful for high-pressure cell sorting as Hepes controls pH better at high pressure than phosphate buffers do.

Finally, since the sheath and sample core stream do not mix, you can use water as a sheath fluid on analyzers. Adding a small amount (0.1%) of 2-phenoxyethanol will help as this serves as a surfactant, helping keep the system flowing by reducing the surface tension.

5. Sample Injection Port

In flow cytometry, suspended cells are moved through the flow cytometer’s tubing all the way to the interrogation point and finally into the waste (or to be sorted and recovered). To do this, the fluidics components of the flow cytometry are required.

The fluidics are comprised of three components. The first component is a running fluid (or sheath fluid – see above), that runs through the system in laminar flow. The movement of this sheath can be achieved by several mechanisms, the most common method using pressure provided by pumps.

The second component of the fluidics is the sample injection port (SIP). This is where the sample is pushed through the tubing to be introduced to the sheath fluid. Based on the principles of hydrodynamic focusing, these cells are strung out, single file, in the direction of the flow, where they will pass the interrogation point.

The third and final main component of the fluidics is the flow cell, which is where the first two components and the suspended cells themselves come together.

6. Isotype Control

The“ isotype” in isotype control refers to the genetic variation in the heavy and light chains that make up the whole antibody moiety. In mammals, there are 9 possible heavy chain isotypes and two light chain isotypes. Every antibody will have a specific isotype, and this is available on the technical information spec sheet.

For example, you might have an antibody with an isotype of IgG1, kappa. This indicates the heavy chain is of the IgG1 isotype. Where things get interesting is that these isotypes can have different non-specific binding affinity to cells, which has lead to investigators using isotype controls, as a control to identify where cells are positive or negative.

The issue with isotype controls is that they are not proper gating controls. Instead, these controls should only be used for identifying potential blocking problems. For more on this, read the following paper: Herzenberg, LA et al. Interpreting flow cytometry data.

7. Antibody Titration

Titration is the process of identifying the best concentration to use an antibody for a given assay. While the antibody’s vendor will provide a specific concentration to use, this may not be appropriate for your assay.

Performing titration is a simple process: fix the cell concentration, the time of incubation, the volume of reaction, and temperature. The graph below displays an antibody that was used to stain 1×106 cells for 20 minutes on ice. To identify the best concentration to use, the modified Staining Index (SI) was calculated and plotted against the concentration. As is shown by this figure, as the concentration increases above 0.5 μg/ml, the SI decreases, due in part to the increase in the background (non-specific staining).

Screen Shot 2016-05-03 at 8.01.18 PM

At concentrations below 0.25  μg/ml, the SI decreases because the antibody is no longer at a saturating concentration. Therefore, the best concentration to use is between 0.25-0.5 μg/ml. Titration helps save money and reagents, ensures the optimal concentration of reagent is being used, and avoids background due to high concentration of antibodies.

8. Differential Pressure

Differential pressure-based flow cytometers currently dominate the market. These systems have two pressure regulators. The first is at a constant pressure that sets how fast the fluids run through the system. The second is regulated by you, the scientist.

As the sample pressure goes from low, to medium, to high, the pressure on the sample increases. This results in the volume of the sample increasing (from ~15 ml/min to ~60 ml/min). The difference between the sample pressure and the sheath pressure is the differential pressure. This controls the width of the core stream and the total number of cells passing the laser intercept point.

9. Jablonski Diagram

The Jablonski diagram illustrates the electronic states of a molecule as well as the transitions between them. These states are arranged vertically by energy, and grouped horizontally by spin multiplicity.

Flow Cytometry Terms And Definitions | Expert Cytometry | flow cytometry meanings

In the below image, nonradiative transitions are indicated by straight arrows and radiative transitions by squiggly arrows. The vibrational states of each electronic state are indicated with parallel horizontal lines. For flow cytometry, it is important to note that the energy of the emission is usually less than that of the absorption. As such, fluorescence normally occurs at lower energies or longer wavelengths.

10. Bandpass, Shortpass, And Longpass Filters

A bandpass filter is a filter that allows light between a set wavelength to pass through it, reflecting only light above and below the set wavelength. For example, a bandpass filter with a wavelength of 550/40nm would allow light between 530nm and 570nm to pass through, but reflect light below 530nm and above 570nm.

A shortpass filter is a filter that allows light over a set wavelength to pass through and reflects light above the set wavelength. For example, a shortpass filter with a wavelength of 450nm would allow light with a wavelength less than 450nm to pass through the filter, but reflect light higher than 450nm.

A longpass filter, on the other hand, is a filter that allows light over a set wavelength to pass through and reflects light below the set wavelength. For example, a longpass filter with a wavelength of 670nm would allow light with a wavelength greater than 670nm to pass through the filter, but reflect light lower than 670nm.

11. Spectral Profile And Spectral Viewer

Every fluorophore has a unique excitation and emission profile which is usually displayed on a spectral viewer, or spectral graph. The combination of the excitation and emission profiles is the fluorophore’s spectral profile. Every fluorophore has a peak excitation wavelength (the wavelength at optimal excitation) and a peak emission wavelength (the wavelength of optimal detection). Each fluorophore will also have a much larger range of excitation and emission wavelengths at reduced optimization. This “curve” is what is displayed on a spectral viewer.

The spectral profile of a fluorophore is used to determine the excitation and detection efficiency at any given wavelength. The spectral profile aids in panel design and selecting optimal fluorophores for a given instrument. The spectral profile can also help in determining compensation considerations. There are numerous resources available to view the spectral profiles of various fluorophores, including this resource from eBioscience.

12. FACS Analysis

Flow cytometry is the science of measuring the physical and biochemical processes on cells and cell-like particles. This analysis is performed in an instrument called the flow cytometer.  FACS Analysis is the shorthand expression for this type of cell analysis. The term FACS stands for Fluorescent Activated Cell Sorting, a term first coined by Len Herzenberg in the 1970’s, and later trademarked by Becton Dickinson. Since that time, FACS has come to be used as a generic term for all of flow cytometry, even though it is a specific trademarked term.

Understanding the above terms, as well as the proper definition of each term, will help you perform better flow cytometry experiments. It will also help you intelligently communicate your methodologies and results in grant submissions and peer-reviewed paper submissions. By being careful not to misuse words such as FACS Analysis, or misunderstand words such as logicle scaling, you will be seen as competent in the field of flow cytometry and your grants and paper will stand better odds of being awarded and passing the dreaded third reviewer, respectively.

To learn more about getting your flow cytometry data published and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class

ABOUT TIM BUSHNELL, PHD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Tim Bushnell, PhD

Similar Articles

The Power Of Spectral Viewers And Their Use In Full Spectrum Flow Cytometry

The Power Of Spectral Viewers And Their Use In Full Spectrum Flow Cytometry

By: Tim Bushnell, PhD

What photon from yonder fluorochrome breaks?  It is … umm… hmmm. Let me see. Excitation off a 561 nm laser, with an emission maximum of 692 nm. I’m sure if Shakespeare was a flow cytometrist, he might have written that very scene. But the play is lost in time. However, since the protagonist had difficulty determining what fluorochrome was emitting photons, let’s consider how this could be figured out. In my opinion, one of the handiest flow cytometry tools is the spectral viewer. This tool helps visualize the excitation and emission profile of different fluorochromes, as well as allowing you…

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

By: Tim Bushnell, PhD

Reproducibility has been an ongoing, and important, concept in the sciences for years.  In the area of biomedical research, the alarm was sounded by several papers published in the early 2010’s.  Authors like Begley and Ellis, Prinz and coworkers, and Vasilevsky and colleagues, among others reported an alarming trend in the reproducibility of pre-clinical data.  These reports indicated between 50% to almost 90% of published pre-clinical data were not reproducible.  This was further highlighted in the article by Freedman and coworkers, who tried to identify and quantify the different sources of error that could be causing this crisis.  Figure 1,…

5 Common Flow Cytometry Questions, Answered

5 Common Flow Cytometry Questions, Answered

By: Tim Bushnell, PhD

I want to thank all of you who send us your questions about flow cytometry, so I thought I would dip into the old email bag and answer a few of the common ones here.  If your question isn’t answered this time, look for it to be answered in a future blog post.  Of course, if you want us to cover a specific topic, drop us a line.  1. How Fast Can I Go? This is  a common question. The allure of the ‘hi’ button is hard to resist.  The faster you go, the sooner you are finished with data…

Combining Flow Cytometry With Plant Science, Microorganisms, And The Environment

Combining Flow Cytometry With Plant Science, Microorganisms, And The Environment

By: Tim Bushnell, PhD

My first introduction to flow cytometry was talking to a professor who’d brought one on a research cruise to study phytoplankton. It was only later that I was introduced to the marvelous world that’s been my career for over 20 years.   In that time, I’ve had the opportunity to work with researchers in many different areas, exposing me to a wide variety of cell types and more important assays. What continues to amaze me is the number of different parameters we can measure, not just the number of fluorochromes, but the information we can extract from samples – animal, vegetable…

Common Numbers-Based Questions I Get As A Flow Cytometry Core Manager And How To Answer Them

Common Numbers-Based Questions I Get As A Flow Cytometry Core Manager And How To Answer Them

By: Tim Bushnell, PhD

Numbers are all around us.  My personal favorite is ≅1.618 aka ɸ aka ‘the golden ratio’.  It’s found throughout history, where it has influenced architects and artists. We see it in nature, in plants, and it is used in movies to frame shots. It can be approximated by the Fibonacci sequence (another math favorite of mine). However, I have not worked out how to apply this to flow cytometry.  That doesn’t mean numbers aren’t important in flow cytometry. They are central to everything we do, and in this blog, I’m going to flit around numbers-based questions that I have received…

3 Must-Have High-Dimensional Flow Cytometry Controls

3 Must-Have High-Dimensional Flow Cytometry Controls

By: Tim Bushnell, PhD

Developments such as the recent upgrade to the Cytobank analysis platform and the creation of new packages such as Immunocluster are reducing the computational expertise needed to work with high-dimensional flow cytometry datasets. Whether you are a researcher in academia, industry, or government, you may want to take advantage of the reduced barrier to entry to apply high-dimensional flow cytometry in your work. However, you’ll need the right experimental design to access the new transformative insights available through these approaches and avoid wasting the considerable time and money required for performing them. As with all experiments, a good design begins…

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

By: Tim Bushnell, PhD

Fluorochrome, antibodies and detectors are important. The journey of a thousand cells starts with a good fluorescent panel. The polychromatic panel is the combination of antibodies and fluorochromes. These will be used during the experiment to answer the biological question of interest. When you only need a few targets, the creation of the panel is relatively straightforward. It’s only when you start to get into more complex panels with multiple fluorochromes that overlap in excitation and emission gets more interesting.  FLUOROCHROMES Both full spectrum and traditional fluorescent flow cytometry rely on measuring the emission of the fluorochromes that are attached…

Flow Cytometry Year in Review: Key Changes To Know

Flow Cytometry Year in Review: Key Changes To Know

By: Meerambika Mishra

Here we are, at the end of an eventful year 2021. But with the promise of a new year 2022 to come. It has been a long year, filled with ups and downs. It is always good to reflect on the past year as we move to the future.  In Memoriam Sir Isaac Newton wrote “If I have seen further, it is by standing upon the shoulders of giants.” In the past year, we have lost some giants of our field including Zbigniew Darzynkiwicz, who contributed much in the areas of cell cycle analysis and apoptosis. Howard Shapiro, known for…

What Star Trek Taught Me About Flow Cytometry

What Star Trek Taught Me About Flow Cytometry

By: Tim Bushnell, PhD

It is no secret that I am a very big fan of the Star Trek franchise. There are many good episodes and lessons explored in the 813+ episodes, 12 movies (and counting). Don’t worry, this blog is not going to review all 813, or even 5 of them. Instead, some of the lessons I have taken away from the show that have applicability to science and flow cytometry.  “Darmok and Jalad at Tanagra.”  (ST:TNG season 5, episode 2) This is probably one of my favorite episodes, which involves Picard and an alien trying to establish a common ground and learn…

Top Industry Career eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.